Specifications
| SANYO Sterisonic Gxp Series | |||
| Single Chamber Model Number | MCO-19AIC-UVH | MCO-19AIC-UV | MCO-19AIC |
| Dual Chamber Model Number | MCO-38AIC-UVH | MCO-38AIC-UV | MCO-38AIC |
| Major Operating Systems | |||
| H2O2 Sterilization System | Standard | Optional | Optional |
| SafeCell™ UV System | Standard | Standard | Optional |
| Single Beam, Dual Detector IR CO2 Sensor | Standard | Standard | Standard |
| InCuSaFe™ Copper Enriched Stainless Steel Interior | Standard | Standard | Standard |
| LCD Graphical Controller/Display, Door Mounted | Standard | Standard | Standard |
| Direct Heat, Air™ (DHA) Air Jacket | Standard | Standard | Standard |
| Sterilization and Decontamination | |||
| H2O2 Sterilization System | Vaporization in situ | Optional | Optional |
| Interior UV Lamp, Programmable, Ozone Free | Standard | Standard | Optional |
| Copper Enriched Stainless Steel Interior with Germicidal Protection | Standard | Standard | Standard |
| Environmental Performance | |||
| Temperature Control Range | +5°C above ambient to 50°C (in a 5°C to 35°C ambient). | ||
| Temperature Control Uniformity and Deviation | ±0.25°C in 25°C ambient, setting 37°C, 5% CO2, no load | ||
| CO2 Control Range and Deviation | 0% to 20%, ±0.15% in 25°C ambient, setting 37°C, 5% CO2, no load | ||
| CO2 Sensor Platform | Ceramic based, single beam, dual wavelength measurement of actual vs. contrast, with continuous auto-zero calibration. | ||
| CO2 Sampling Patent Pending | No moving parts; airflow passes over in/out ports to sustain Venturi effect, continuous sampling | ||
| CO2 Calibration | Single point zero automatic; semi-automatic one point span @ reference optional. | ||
| Airfl ow | Gentle vertical airfl ow, continuous with inner door closed. | ||
| Interior Humidity | 95%RH @ 37°C through evaporation via DHA™ heating system; reflective optical low water sensor. | ||
| Control, Monitoring, Alarm | |||
| Temperature and CO2 Control | P.I.D, setpoint resolution 0.1°, 0.1% | ||
| Display | Alphanumeric LCD digital display messaging. | ||
| Data Acquisition | Automatic log function of temperature and CO2. | ||
| Communications | Remote alarm contacts standard. Optional 4-20mA connection. Optional PC interface, Catalog No. MTR-480 with RS232/RS485 data ports available. |
||
| Cabinet Design and Construction | |||
| Superstructure, Exterior Cabinet and Door | Galvanized steel exterior, baked-on enamel finish | ||
| Interior and Shelves | Copper-enriched stainless steel | ||
| Inner Door | Tempered glass | ||
| Insulation | Rigid foam polyurethane | ||
| Outer Door | Reversible, heated. | ||
| Access Port | Single opening with interior and exterior 30mm non-VOC silicone stoppers. | ||
| Leveling Feet | 4, Adjustable. | ||
| Energy, Electrical, Utilities | |||
| Maximum Power Consumption | 310W | 310W | 310W |
| Maximum Heat Discharge | 1062 BTU | 1062 BTU | 1062 BTU |
| Electrical | 115V,60Hz with NEMA 5-15 plug provided; requires NEMA 5-15R grounded receptacle. | ||
| CO2 Gas Connection | 4 to 6mm inner diameter tubing | ||
| CO2 Gas Pressure | Nominal 4.3 PSI from two-stage CO2 regulator | ||
| CO2 Gas Cylinder Switchover System | Optional | Optional | Optional |
| Dimensions, Weights, Capacities | |||
| Interior | 19.3" W x 20.6" F-B x 26.2" H (490 x 523 x 665 mm) | ||
| Exterior | 24.4" W x 27.9" F-B x 35.4" H (620 x 710 x 900 mm) | ||
| Volume | 6.0 cu.ft. (170 liters) | ||
| Shelves | Maximum 15/chamber, 4 supplied standard; 17" x 17" (450mm), 15.4 lbs/7kg capacity | ||
| Net Weight | 205 lbs (93kg) nominal | ||