Sterisonic™ GxP Design
Elevated Humidity, Low Water Level Warning
To avoid cell culture desiccation, the SANYO Sterisonic™ GxP CO2 incubator maintains ~95% RH at 37°C.
- Humidification is achieved by a combined forced-air and natural evaporation method enhanced by the Direct Heat and Air Jacket™ base heater, and protected by an optical water level indicator to warn of low water in the removable humidity pan.
- The humidity removes easily for regular cleaning and refill.
- When filled the pan slides into place and the optical sensor returns to position and the SafeCell™ UV lamp destroys any contaminants introduced during the process.
- For the H2O2 in situ sterilization cycle the humidity pan is repositioned against the chamber wall to eliminate the need for separate autoclaving.
Ergonomic Cabinet Design
With reversible inner and outer doors, a single SANYO incubator offers the industry's most flexible installation options available.
- Low profile cabinet with door-mounted control panel permits easy access and viewing.
- The outer door latch and door heater cable is easily switched if a reverse opening is required.
- Cabinet knock-outs are pre-drilled and tapped to eliminate drilling and to simplify re-mounting of door hardware.
- The outer door closes against a soft, easily cleaned magnetic gasket designed to eliminate ambient air shear across the glass inner door, minimizing condensation.
- A door ajar alarm provides an audible and visual warning if the outer door is left open.
- A center-mounted key lock is located beneath the door for added security; model MCO-19AIC(UVH) only.
Field Reversible Door
The field-reversible door allows universal installation using the left-hand hinge (standard) or a right-hand hinge modification
- The outer door includes a universal finger grip at each side.
- Mounting holes for hinge hardware are pre-drilled and capped with easily removable trim plugs.
- The door heater cable plugs into the alternate connection to complete the change.
Inner Door and Gasket
The inner design is critical to successful contamination control technique.
- The inner gasket body forms an effective thermal transition between the ambient air and warm, humidified incubator atmosphere, minimizing condensation and eliminating moisture traps which can harbor contaminants.
- The inner door gasket is a dual durometer extrusion (two levels of softness) from closed-cell silicone to inhibit contamination.
- The gasket feather-edge allows the inner glass door to close gently against the chamber opening for a tight peripheral seal.
- The entire inner door gasket is removable for cleaning and/or replacement if required.
- Radiant heat from the outer door, apportioned by the microprocessor control as a function of the patented Direct Heat and Air Jacket™, automatically warms the inner door glass in proportion to total heat demand and condensation control.